- The Konica Minolta Chroma Meter CL-200A elevates lighting measurements with its compact design. Ideal for various light sources like LED and OLED, it offers versatile measurements, including illuminance, chromaticity, correlated color temperature, tristimulus values, dominant wavelength, purity of light, and more. With a detachable measuring head and multi-point options, it is a valuable tool for various applications like projector light source research, LED billboard development, lighting research, and more.
Chroma Meter CL-200A
The Konica Minolta Chroma Meter CL-200A, successor to the CL-200, is designed for the evaluation of various light sources, including LED, Organic LED (OLED), etc. This versatile lux meter is capable of measuring illuminance, chromaticity, correlated color temperature, tristimulus values, dominant wavelength, purity of light, and more.
The Chroma Meter CL-200A conforms to Class AA of JIS C 1609-1 standards, offering unmatched quality and precision. It comes equipped with a removable receptor head, which can be used in darkroom boxes or installed in an inspection system. Data is collected by connecting the receptor and body via a LAN cable. The Chroma Meter CL-200A also features the ability for multi-point measurements, operating up to 30 receptor heads, along with the user-friendly calibration function. The calibration function provides the ability to compare measured values to a specific sample.
Its compact, lightweight design is created to fit comfortably in the palm of your hand and for easy portability. The inclusion of the CL-S10w Data Management SoftwareCL-S10w Data Management Software, which includes functions like LED ranking, JIS correlated color temperature, user calibration, etc., is provided with the Chroma Meter CL-200A to facilitate easy data transfer and seamless computer connectivity.
The Chroma Meter CL-200A is your go-to tool for a multitude of applications, from projector light source research and LED billboard development to light distribution evaluation and building and interior lighting research.
Chroma Meter CL-200A
- Compact – hand-held, lightweight design
- Portable – battery operated, receptor head can also be detached from the main body for remote measurements
- User-friendly – low battery indicator, can be controlled by a PC, easily transfer data to Excel, and includes a calibration function
- Simultaneous measurements – has the ability to determine illuminance, color temperature, chromaticity, dominant wavelength, and excitation purity of a light source
- Ideal for white LEDs – can measure the chromaticity from the phosphor as well as inspect the final product
- Multi-use – can be used for any light source, ideal for LEDs and OLEDs
- LED ranking function – using the included software, users can rank measurements on a chromaticity chart for easy comparison
Chroma Meter CL-200A
Relative Spectral Response Closely matches CIE Standard Observer curves X(λ), y(λ), and Z(λ) Within 6%(f1’) of the CIE spectral luminous effeciency V(λ) Cosine response(f2) Ev: Within 3% Receptor Silicon photocell Measuring function Tristimulus values:XYZ
Chromaticity : Ev xy, Ev u’v’
Correlated color temperature: Ev, Tcp, Δuv
Color difference: Δ(XYZ), Δ(Ev, xy), Δ(Ev u’v’), ΔEv Δu’v’Other function User calibration function, Data hold function,
Multi-point measurement(2 to 30 points)Measuring range 0.1∼99,990lx, 0.01∼9,999 fcd(Chromaticity:5 lx, 0.5 fcd or above) in four automatically selected ranges(lx or fcd is swichable) Accuracy Ev: ±2%±1digit of displayed value
xy: ±0.002(800lx, standardilluminant A measured)Repeatability Ev: 0.5%+1 digit(2σ)
xy: ±0.0005Temperature drift Ev: ±3%±1digit of displayed value,
xy: ±0.003Humidity drift Ev: ±3%±1digit of displayed value,
xy: ±0.003Response time 0.5 seconds (each measurement) Digital output USB 2.0 full-speed Display 4 Significant-digit LCD with back-light illumination Operating temperature/ Humidity range -10 to 40°C, relative humidity 85% or less
(at 35°C) with no condensationStorage temperature/ Humidity range -20 to 55°C, relative humidity 85% or less
(at 35°C) with no condensationPower source 2 AA-size batteries/AC adapter (optional) Battery life 72 hours or longer (when alkaline bateries are used) in continuous measurement Dimensions 69x174x35mm Weight 215g(7.6 oz.) not including batteries Standard accessories Case T-A10, Cap T-A13, Neck Strap, batteries, Data Management Software CL-S10w, Connecting Cable T-A15 Optional accessories Recepter Head, Adapter Unit for Main Body T-A20,
Adapter Unit for Receptor Head T-A21,
AC Adapter AC-308(available for 1 to 10 Receptor Heads)
AC Adapter AC-311(available for 1 to 30 Receptor Heads)
Printer Cable T-A12, Hood CL-A11, Case CL-A10Data Management Software CL-S107w Specification
Software type MicroSoft Excel add-in software Excel is required for operation of this software Operating Systems Windows XP + Excel 2003 (Japanese, English,Chinese)
Windows 7 + Excel 2003 (Japanese, English, Chinese)Hardware to connect CL-200A CL-200 some of the functions are not available when connected with the CL-200 ∗Specfications are subject to change without notice.
∗Company names and product names mentioned here are the registered trademarks or trademarks of their respective caompanies.